Amplifiers / Isolators      Drug Application      Modules      Data Acquisition      Signal Processing      Temperature Controllers      Micropositioners      
       Perfusion Systems      Miscellaneous Products      

















 
 


The VA-10X is a sensitive (picoampere range) current amplifier that is intended for voltammetric measurements with carbon-fiber microelectrodes in biological systems, where the total currents do not exceed a few tens of nA.
It was designed at the Max-Planck-Institute for Experimental Medicine in Göttingen as an economically priced alternative to do-it-yourself systems and expensive commercial systems.
It can be used for either DC amperometry using the built-in voltage source, or it can be operated with user-supplied external voltage waveforms (e.g. for cyclic voltammetry).
The VA-10X is ideally suited for measurements from single cells plated onto glass cover slips and with carbon-fiber disk electrodes having diameters of 10 µM or less. npi recommends ALA CFE-1 carbon fiber electrodes. However, it can also be used for measurements made on superficially located cells in tissue slices.
The standard VA-10X is not recommended for use in in-vivo recordings with carbon-fiber electrodes having long cylindrical measuring surfaces, because in this case currents approach the µA range. Ask npi electronic for headstages with different feedback resistors.
Optionally, the VA-10X can be equipped with a frequency booster in order to enhance the bandwidth to several kHz.

The VA-10 is also available as module for the EPMS-07 system.

Download VA-10X manual 249 kB.

Download VA-10 brochure 84 kB.

Download a reference list about VA-10 systems.






Click to enlarge

Download page as PDF





       © copyright 2003 npi electronic - Last update: 15.09.2014 Go to top | Masthead / Disclaimer